Pump Replacemente
Pump Replacemente
Client complaint: No water was flowing during the bath refill procedure.
Initial assessment: After inspecting the system, I identified that water was only reaching the syringe module, which is powered by a gear pump. All other areas of the system were dry.
Diagnosis: The main inlet pump had failed due to cavitation, which caused a blockage and soot accumulation in the tubing.
Action taken:
Replaced the main pump
Flushed the entire tubing system
Confirmed water flow was restored across all modules